3-Series CO-10000 Sensor
DataSheet Download
Product Description

The sensor is designed for the measurement of CO concentration in gas phase. It can be used as the pin to pin replacement of the standard 3 series electrochemical CO sensor.

Performance Characteristics
Nominal Range 0~10000 ppm
Maximum Overload 20000 ppm
Sensitivity(20℃) 0.030 ± 0.006 μA/ppm
Response Time (T90) ≤ 30 s
Zero Signal(20℃) -3 ~ +10 ppm
Baseline Shift (- 20℃ ~ 50℃) < 20 ppm
Resolution 1 ppm
Linearity linear up to 10000ppm
Bias Voltage zero

 

Product Dimensions

Environmental
Temperature Range -20℃ ~ 50℃
Pressure Range 1 atm ± 10%
Humidity Range 15% ~ 90% RH non-condensing
Life time
Long time output drift < 2% signal/month
Recommended Storage Temp 10℃ ~ 30℃
Expected Operating Life 3 years in clean air
Storage Life 6 months in original packaging
Warranty 12 months
Intrinsic safety Data
Maximum Current at 20000ppm CO < 1mA
Maximum O/C Voltage 1.3V
Maximum S/C Current < 1.0A
Mechanical
Housing Material ABS
Weight (Nominal) 23g
Orientation None
Installation

Output signals from the sensor pins are different. Inappropriate use of the pins in product design will affect the sensor functionality. Exposure to high concentrations of solvent vapors should be avoided under any condition. Mechanical overstress may cause deformation or cracks of the plastic enclosure of the sensor. If the sensor is used in extreme environmental conditions, please contact SemeaTech for more details.

Cross-sensitivity Data
Gas Concentration (ppm) Output signal (ppm CO equivalent)
Sulfur Dioxide 100 3.4
Nitrogen Dioxide 90 2.8
Nitric Oxide 250 1.4
Hydrogen 200 121.4
Hydrogen Chloride 100 2.2
Hydrogen Sulfide 200 1.8
Ethylene 200 10

Note: The cross sensitivity are including but not limited to the above gases. It may also respond to other gases.The data in the table above may vary from different batches of sensors and the changes of test environment. Calibration with cross sensitivity gas is not recommended.

Temperature Data

Safety Note:

The sensor is designed to be used in certain instruments for life critical applications. To ensure the sensor functioning per its specifications inside the instrument, it is required to read the instrument user's guide carefully and comply with the calibration procedures by using certified target calibration gas before each use. Failure to do so may cause serious injury and fatality. Please do not open the housing because the electrolyte stored inside is harmful.

It is highly recommended for customers to validate the sensor performance using this document as a reference for their product designs or applications.